A dental practice runs on a small oil-free piston compressor, 1 to 5 hp with two or three heads on a tank, delivering about 2 SCFM for every operatory in use at once: the high-speed handpiece turbine at 30 to 40 psig, the air-water syringe, the chair controls, and the lab's sandblaster and model trimmer. A six-chair office draws 10 to 15 SCFM. A hospital is a different scale, with a duplex medical-air plant under NFPA 99 for patient breathing air, a separate instrument-air system for surgical tools, and then all the air that is not medical gas: laboratory bench air, sterile-processing utility air, the maintenance shop and the pneumatic building controls.
Water in dental air is the complaint the patient notices. Condensate in the handpiece line arrives as a spit of water at the bur, carries the corrosion products of the line with it and shortens turbine bearing life; in the syringe it fogs the mirror. Because dental compressors are oil-free, the contaminant after water is particulate and whatever the intake draws in, which is why the dental filter train ends in a fine coalescing element. A membrane or desiccant dryer is often built into the compressor package; where it is not, or where a clinic with 20 to 60 chairs outgrows it, a small refrigerated dryer with a filter pair is the usual answer.
A refrigerated dryer holds a 38°F pressure dew point at tank pressure, which is roughly 20°F by the time the regulator has dropped the air to handpiece pressure, well below any operatory. It sits after the tank and ahead of the regulator, with a 1 micron prefilter, a 0.01 micron coalescing afterfilter, a wall bracket and a 1701 bypass valve so it can be serviced without closing the practice. The HPRB Series at 7 to 25 SCFM and the DXR Series at 10 to 30 SCFM cover offices; a 30-chair clinic on a 15 hp oil-free scroll uses the 50 to 65 SCFM sizes. Confirm the dew point the practice's authority requires before choosing between a refrigerated dryer and a membrane or desiccant unit.
Hospital medical air is not a refrigerated-dryer application, and this site does not pretend otherwise. NFPA 99 requires the medical-air dryers to deliver a dew point below the frost point, 32°F at line pressure, with an alarm at 35°F, and instrument air for surgical tools to reach -22°F, both of which take desiccant drying. The refrigerated dryers here serve the rest of the building: laboratory air, sterile-processing utility air, shop air and control air, each on its own small compressor.
Request a dental or medical dryer quote
Medical-air dew point alarm under NFPA 99NFPA 99 wants medical air alarmed at 35°F, a level a refrigerated dryer cannot promise, which is why this page keeps its dryers to dental, lab and utility air.
Where it shows up first
Dental air, lab air and utility air; NFPA 99 systems are separate
Ratings at 100 psig and 100°F inlet. A 1 to 5 hp oil-free dental compressor delivers 4 to 18 SCFM; a clinic or laboratory on a 7.5 to 15 hp oil-free scroll delivers 25 to 60 SCFM.
| SERIES | TYPE | FLOW RANGE | MAX INLET | MAX PRESSURE |
| HPRB Series | Non-Cycling | 7 to 50 SCFM | 120°F | 200 psig |
| DXR Series | Non-Cycling | 10 to 4,200 SCFM | 131°F | 203 psig |
| HRHT Series | High Inlet Temperature | 25 to 100 SCFM | 200°F | — |
| GFN Series Filters | Compressed Air Filter | 6 to 1,500 SCFM | — | 232 psig |
| NGF Series Filters | Compressed Air Filter | 20 to 1,500 SCFM | — | 250 psig |
| 1701 Bypass Valves & Mounting Brackets | Bypass & Mounting | 5 to 35 SCFM | — | 200 psig |
| Snap-Trap & Trip-L-Trap Drains | Condensate Drain | Call for sizing | — | 175 psig |
| PDM Dew Point Monitor | Dew Point Monitor | Call for sizing | — | 145 psig |
A dental compressor lives in a closet or a basement corner, and a refrigerated dryer beside it shares the heat. An air-cooled dryer rated at 100°F ambient loses 12 percent of its capacity at 110°F and trips on high refrigerant pressure not far above that, and a closet with a compressor cycling all day reaches those temperatures by mid-afternoon. Give the closet a louvered door or a transfer grille, and keep the dryer's condenser clear of the compressor's exhaust air. Where the closet cannot be ventilated, pipe the compressor to the tank first and the tank to the dryer, so the tank sheds heat before the air reaches the dryer inlet.
Compressor, tank, 1 micron prefilter, dryer, 0.01 micron coalescer, then the regulator and the manifold to the operatories. Drying at tank pressure ahead of the regulator gives the lowest dew point at the chair and keeps the dryer at the pressure it is rated for. Most dental compressors are in a closet or a basement mechanical room, so the dryer goes on a wall bracket beside the tank with a 1701 bypass valve, and it needs a few inches of clearance for its condenser and a room that stays below about 100°F. The lab and dental compressor page walks through the installation.
Figure 2 SCFM per operatory in use at the same time, so a six-chair office peaks at 10 to 12 SCFM and a 15 SCFM dryer covers it, while a twelve-chair clinic on a 5 hp compressor takes the 25 SCFM size. Size to the compressor's delivery when it is larger than the chair count suggests.
NFPA 99 governs the piped medical gas systems of a health care facility. Its medical-air plant must be duplex, oil-free and dried below the frost point with a dew point alarm at 35°F, and its instrument air must reach -22°F, so both run on desiccant dryers and are engineered and certified as medical gas systems. It does not turn every compressor in the building into a medical gas source: the laboratory, the sterile-processing department, the maintenance shop and the pneumatic building controls run on their own compressors, and those are the systems a refrigerated dryer and a filter pair serve. Dental air answers to the practice's own authority, and the drying requirement varies, so check it before choosing the dryer type.
An oil-free piston or scroll compressor removes the oil aerosol but not the wear particles, the intake dust or the moisture, and a handpiece turbine is no more tolerant of grit than of water. Keep the 1 micron prefilter and the 0.01 micron coalescing afterfilter, change them on the annual service, and drain the tank with a zero-loss drain rather than a valve someone is suppose to open weekly.
Only if the closet is ventilated or the dryer is sized for the heat. An air-cooled unit is rated at 100°F ambient and gives up about 12 percent of its capacity at 110°F, and a closet holding a compressor that cycles all day gets there easily. The usual symptom is a dryer that trips on high refrigerant pressure late in the afternoon, resets overnight and looks fine to the technician the next morning. A louvered door or transfer grille is the cheapest fix.
Water at the handpiece, or a fogging mirror, late in a busy day when most operatories are running together. Allow roughly 2 SCFM per chair in simultaneous use and compare the total with the rating of the membrane or desiccant unit built into the compressor package. If the new chair count is beyond it, a small refrigerated dryer with a prefilter and a coalescing afterfilter, mounted at tank pressure ahead of the regulator, takes over the drying duty.
Everything outside the NFPA 99 piped medical gas systems. That covers laboratory bench air, sterile-processing utility air for washer and sterilizer valves, the maintenance shop and the pneumatic building controls, each normally on its own compressor. Medical air and surgical instrument air are separate certified plants on desiccant dryers, because their dew points are beyond any refrigeration circuit. Keep those systems physically apart; tying a utility compressor into a medical gas line is a code problem, not a sizing one.
The oil-free compressor packages behind dental offices and laboratories, matched to the 7 to 50 SCFM dryers that fit.
Learn More
Fit a 1 micron prefilter and a 0.01 micron afterfilter, and put the dryer where sawdust cannot reach its condenser.
Learn More
A mill in a hot climate with tower water at the compressors usually takes the water-cooled version, which holds its rating on the hottest day.
Learn More